Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
944
User Manual 1178.8520.02 ─ 08
Suffix:
<sb>
.
1 to 8
<ri>
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
ESP:RANG3:SWE:TIME:AUTO OFF
Deactivates the sweep time auto mode for range 3.
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:TRANsducer
<Transducer>
This command selects a transducer factor for a SEM range.
Note that
●
the transducer must cover at least the span of the range
●
the x-axis has to be linear
●
the unit has to be dB
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the measurement range.
Parameters:
<Transducer>
String containing the transducer file name, including the path
information.
Manual operation:
See
[SENSe:]ESPectrum<sb>:SSETup
<State>
Enables or disables symmetrical configuration of the range settings.
See
Chapter 5.6.4.1, "Ranges and range settings"
Suffix:
<sb>
.
1 to 8
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...