c) Select "Upload" to copy the file from the PC to the defined folder on the
R&S
FSV/A.
11.7.6
How to deactivate the web browser interface
If you want to prevent other users in the LAN from accessing or operating the
R&S
FSV/A via its LAN web browser interface, you must deactivate this function. Note
that
after a firmware update
the function is
automatically active
again until you
deactivate it manually.
To deactivate the LAN web browser interface
1.
Select the "Windows" icon in the toolbar to access the operating system.
2. In the "Start" menu, select "Control Panel".
3. Select "System and Security" > "Administrative Tools".
4. From the list on the right, select "Services".
5. From the list of local services, select "R&S TightVNC Server".
6. Set "Startup type" to "Disabled".
7. Select "Stop".
8. Select "Apply".
The next time a user enters the IP address of the instrument in a web browser, an
error message is displayed:
Failed to connect to server (code. 1006)
How to set up a network and remote control