
Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
271
User Manual 1178.8520.02 ─ 08
tions of the absolute power from the limit line for each range, and the results can be
checked against defined limits automatically.
5.7.2
Spurious emissions measurement results
The measured signal, including any spurious emissions, and optionally the detected
peaks are displayed in the Spurious Emissions measurement diagram. If defined, the
limit lines and the limit check results are also indicated. In addition to the graphical
results, a result table can be displayed to evaluate the measured powers and limit
check results (see also
Chapter 5.7.3.2, "Limit lines in spurious measurements"
on page 273). The details of the evaluation list can be configured.
The following information is provided in the evaluation list for each range:
Column
Description
Range Low
Frequency range start for the range the peak value belongs to
Range Up
Frequency range end for the range the peak value belongs to
RBW
RBW of the range
Frequency
Frequency at the peak value
Power Abs
Absolute power level at the peak value
Δ
Limit
Deviation of the absolute power level from the defined limit for the peak
value
By default, one peak per range is displayed. However, you can change the settings to:
●
Display all peaks
Spurious emissions measurement
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...