Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1194
User Manual 1178.8520.02 ─ 08
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:NEXT
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes all frames. It does not change the horizontal position of the
marker.
Suffix:
<n>
.
<m>
Manual operation:
See
"Search Mode for Next Peak in Y-Direction"
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum[:PEAK]
This command moves a marker vertically to the highest level for the current frequency.
The search includes all frames. It does not change the horizontal position of the
marker.
If the marker hasn't been active yet, the command looks for the peak level in the whole
spectrogram.
Suffix:
<n>
.
<m>
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:ABOVe
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:ABOVe
This command moves a marker vertically to the next higher minimum level for the cur-
rent frequency.
The search includes only frames above the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
<m>
Manual operation:
See
"Search Mode for Next Peak in Y-Direction"
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:BELow
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:BELow
This command moves a marker vertically to the next higher minimum level for the cur-
rent frequency.
The search includes only frames below the current marker position. It does not change
the horizontal position of the marker.
Analyzing measurements (basics)
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...