Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1105
User Manual 1178.8520.02 ─ 08
Parameters:
<Span>
The minimum span for measurements in the frequency domain
is 10
Hz.
Range:
0 Hz to fmax
*RST:
Full span
Default unit: Hz
Manual operation:
See
See
See
See
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Manual operation:
See
[SENSe:]FREQuency:STARt
<Frequency>
This command defines a start frequency for measurements in the frequency domain.
Parameters:
<Frequency>
0 to (fmax - min span)
*RST:
0
Default unit: HZ
Example:
FREQ:STAR 20MHz
Manual operation:
See
See
[SENSe:]FREQuency:STOP
<Frequency>
This command defines a stop frequency for measurements in the frequency domain.
Parameters:
<Frequency>
min span to fmax
*RST:
fmax
Default unit: HZ
Example:
FREQ:STOP 2000 MHz
Manual operation:
See
See
12.7.2.2
Configuring signal tracking
When signal tracking is activated, the maximum signal is determined after each fre-
quency sweep and the center frequency is set to the frequency of this signal. Thus with
drifting signals the center frequency follows the signal.
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...