Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1072
User Manual 1178.8520.02 ─ 08
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see the R&S
FSV/A I/Q Analyzer and I/Q Input User Manual.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<CMOffset>
Offset of the mean voltage between the positive and negative
input terminal vs. ground
Range:
-16 V to +16 V
Default unit: V
Manual operation:
See
"Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
[SENSe:]PROBe<pb>:SETup:DMOFfset
<DMOffset>
Sets the DM-mode offset. The setting is only available if a modular probe in DM-mode
is connected to the R&S
FSV/A.
If the probe is disconnected, the DM-mode offset of the probe is reset to 0.0
V.
Note that if the offset for DM-mode or CM-mode is changed, the offsets for the P-mode
and N-mode are adapted accordingly, and vice versa.
For details see the R&S
FSV/A I/Q Analyzer and I/Q Input User Manual.
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<DMOffset>
Voltage offset between the positive and negative input terminal
Default unit: V
Manual operation:
See
"Common Mode Offset / Diff. Mode Offset / P Offset / N Off-
[SENSe:]PROBe<pb>:SETup:MODE
<Mode>
Suffix:
<pb>
.
1..n
Selects the connector:
3 = RF
Parameters:
<Mode>
RSINgle | NOACtion
RSINgle
Run single: starts one data acquisition.
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...