
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1061
User Manual 1178.8520.02 ─ 08
[SENSe:]MIXer<x>:HARMonic:TYPE
<OddEven>
This command specifies whether the harmonic order to be used should be odd, even,
or both.
Which harmonics are supported depends on the mixer type.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<OddEven>
ODD | EVEN | EODD
ODD | EVEN | EODD
*RST:
EVEN
Example:
MIX:HARM:TYPE ODD
Manual operation:
See
[SENSe:]MIXer<x>:HARMonic[:LOW]
<HarmOrder>
This command specifies the harmonic order to be used for the low (first) range.
Suffix:
<x>
.
1..n
irrelevant
Parameters:
<HarmOrder>
Range:
3 to 128 (USER band); for other bands: see band
definition
*RST:
4 (for band U)
Example:
MIX:HARM 3
Manual operation:
See
[SENSe:]MIXer<x>:IF?
Queries the intermediate frequency currently used by the external mixer.
Suffix:
<x>
.
1..n
irrelevant
Example:
MIX:IF?
Example:
See
"Programming example: working with an external mixer"
Usage:
Query only
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...