Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1031
User Manual 1178.8520.02 ─ 08
Example:
LAY:SPL 1,4,70
Moves the splitter between window 1 ('Frequency Sweep') and 3
('"Marker Peak List"') towards the top (70%) of the screen.
The following commands have the exact same effect, as any
combination of windows above and below the splitter moves the
splitter vertically.
LAY:SPL 3,2,70
LAY:SPL 4,1,70
LAY:SPL 2,1,70
Usage:
Setting only
LAYout:WINDow<n>:ADD?
<Direction>,<WindowType>
This command adds a measurement window to the display. Note that with this com-
mand, the suffix <n> determines the existing window next to which the new window is
added. Unlike
, for which the existing window is defined by
a parameter.
To replace an existing window, use the
command.
This command is always used as a query so that you immediately obtain the name of
the new window as a result.
Suffix:
<n>
.
Query parameters:
<Direction>
LEFT | RIGHt | ABOVe | BELow
<WindowType>
Type of measurement window you want to add.
See
on page 1026 for a list of availa-
ble window types.
Return values:
<NewWindowName> When adding a new window, the command returns its name (by
default the same as its number) as a result.
Example:
LAY:WIND1:ADD? LEFT,MTAB
Result:
'2'
Adds a new window named '2' with a marker table to the left of
window 1.
Usage:
Query only
LAYout:WINDow<n>:IDENtify?
This command queries the
name
of a particular display window (indicated by the <n>
suffix) in the active channel.
Note
: to query the
index
of a particular window, use the
command.
Configuring the result display
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...