Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1017
User Manual 1178.8520.02 ─ 08
Return values:
<PowerLevel>
Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with
The result is a list of floating point values separated by commas.
The unit depends on
Usage:
Query only
[SENSe:]LIST:POWer[:SEQuence]
{<Frequency>, <RefLevel>, <RFAttenuation>,
<ElAttenuation>,<FilterType>, <RBW>, <VBW>, <MeasTime>,
<TriggerLevel>}...
This command configures and initiates the List Evaluation measurement.
The list can contain up to 200 entries (frequencies). You can define a different instru-
ment setup for each frequency that is in the list.
If you synchronize the measurement with
*OPC
, the R&S
FSV/A produces a service
request when all frequencies have been measured and the number of individual mea-
surements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see
Parameters:
<Frequency>
Defines the frequency. Each frequency corresponds to one list
entry.
Range:
0 to Fmax
Default unit: Hz
<RefLevel>
Defines the reference level for a list entry.
Range:
-130 to 30
Increment: 0.01
Default unit: dBm
<RFAttenuation>
Defines the RF attenuation for a list entry.
Range:
0 to 70
Increment: 1
Default unit: dB
<RFAttenuation>
numeric value
Defines the electronic attenuation for a list entry.
A setting other than 0 (OFF) requires option R&S
FSV3-B25.
Range:
0 to 30
Increment: 0 (OFF)
Default unit: dB
<FilterType>
Selects the filter type for a list entry. For more information see
[SENSe:]BANDwidth[:RESolution]:TYPE
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...