Remote Commands
R&S
®
FPL1000
813
User Manual 1178.3370.02 ─ 03
Manual operation:
See
" Export Trace to ASCII File "
MMEMory:STORe<n>:SPURious
<FileName>
This command exports the marker peak list available for spurious emission measure-
ments to a file.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:SPUR 'test'
Saves the current marker peak list in the file test.dat.
Usage:
Event
MMEMory:STORe<n>:TRACe
<Trace>, <FileName>
This command exports trace data from the specified window to an ASCII file.
For details on the file format see
Chapter 7.13.4, "Reference: ASCII File Export For-
Suffix:
<n>
.
Setting parameters:
<Trace>
Number of the trace to be stored
(This parameter is ignored if the option "Export all Traces and all
Table Results" is activated in the Export configuration settings,
see
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR1:TRAC 1,'C:\TEST.ASC'
Stores trace 1 from window 1 in the file
TEST.ASC
.
Usage:
Setting only
Manual operation:
See
" Export Trace to ASCII File "
8.9.6 Examples: Managing Data
...................................................................................815
................................................................................. 815
Managing Settings and Results