Remote Control
R&S
®
FPC
332
User Manual 1178.4130.02 ─ 07
Parameters:
<Scaling>
LINear
Selects a linear scale (in %).
LOGarithmic
Selects a logarithmic scale (in dB).
*RST:
LOGarithmic
Example:
//Select scaling of the y-axis
MEAS:MODE REFL
MEAS:FORM MAGN
DISP:MAGN:Y:SPAC LIN
Manual operation:
See
DISPlay<n>:PHASe:REFerence
<RefValue>
This command defines the reference value for the phase measurement format.
Prerequisites for this command
●
Select reflection measurement (
).
●
).
Suffix:
<n>
.
irrelevant
Parameters:
<RefValue>
<numeric value>
Range:
-100000 to 100000
*RST:
0
Default unit: deg
Example:
//Define reference value for phase
MEAS:MODE REFL
MEAS:FORM PHAS
DISP:PHAS:REF 360
Manual operation:
See
DISPlay<n>:PHASe:REFerence:POSition
<RefPosition>
This command defines the position of the reference value for the phase measurement
format.
Each step shifts the reference position one grid line up or down.
Prerequisites for this command
●
Select reflection measurement (
).
●
).
Suffix:
<n>
.
irrelevant
Remote Commands of the VNA Application