Remote Control
R&S
®
FPC
326
User Manual 1178.4130.02 ─ 07
CALCulate<n>:DTF:DISTance:STARt
<StartDistance>
This command defines the starting point (distance) of the DTF measurement.
Prerequisites for this command
●
Select the DTF measurement (
Suffix:
<n>
.
irrelevant
Parameters:
<StartDistance>
<numeric value>
The unit depends on your selection (
*RST:
0 m
Default unit: m or ft
Example:
//Define start distance for DTF measurement
UNIT:LENG MET
MEAS:MODE DTF
CALC:DTF:DIST:STAR 10
Manual operation:
See
CALCulate<n>:DTF:DISTance:STOP
<StopDistance>
This command defines the end point (distance) of the DTF measurement.
Prerequisites for this command
●
Select the DTF measurement (
Suffix:
<n>
.
irrelevant
Parameters:
<StopDistance>
<numeric value>
The unit depends on your selection (
*RST:
50 m
Default unit: m or ft
Example:
//Define stop distance for DTF measurement
UNIT:LENG MET
MEAS:MODE DTF
CALC:DTF:DIST:STOP 100
Manual operation:
See
CALCulate<n>:TRANsform:DTFault:PEAK:COUNt?
<State>
This command queries the number of entries in the DTF list.
Prerequisites for this command
●
Select the DTF measurement (
Remote Commands of the VNA Application