![Rohde & Schwarz EX-IQ-BOX Скачать руководство пользователя страница 324](http://html1.mh-extra.com/html/rohde-and-schwarz/ex-iq-box/ex-iq-box_user-manual_1477982324.webp)
Remote Control Commands
R&S
®
EX-IQ-BOX
309
User Manual 1409.5570.32 ─ 05
Parameters:
<Source>
RIN | INTernal | CRECovery
RIN
CPRI RE test setup with a connected R&S instrument.
The R&S
EX-IQ-BOX synchronizes the CPRI communication link
with the reference signal provided by an R&S instrument. The 10
MHz signal is fed to the BNC connector REF IN on the back of the
R&S
EX-IQ-BOX.
INTernal
CPRI RE test setup, if the R&S
EX-IQ-BOX is working stand-
alone.
The R&S
EX-IQ-BOX uses its internal reference oscillator for syn-
chronization.
CRECovery
CPRI REC test mode.
The reference clock is extracted from the CPRI link. This selection
is mandatory in a CPRI REC test setup, since the R&S
EX-IQ-BOX
operates in CPRI slave mode.
*RST:
RIN
Example:
SOUR1:EBOX:CPRI:HW:RCL:SOUR RIN
set the internal clock source for reference.
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:HW:SEED:RECeived?
Queries the scrambling seed used by the connected transmitter.
Return values:
<Received>
string
Example:
EBOX:CPRI:HW:SEED:REC?
Response:
'3fffffff'
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:EBOX:CPRI:HW:SEED:SCRambling
<Scrambling>
Specifies the scrambling seed. This parameter applies to CPRI protocol versoin 2.0
Parameters:
<Scrambling>
integer
Range:
depends on other settings to depends on other set-
tings
Example:
EBOX:CPRI:HW:SEED:SCR #H3FFFFFFF
Manual operation:
See
CPRI Remote Control Commands