Remote Control
R&S
®
ESRP
802
User Manual 1175.7445.02 ─ 08
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sequences) and waits for
the end.
Manual operation:
See
[SENSe:]SWEep:COUNt
<NumberSweeps>
This command defines the number of sweeps started with single sweep, which are
used for calculating the average or maximum value. If the values 0 or 1 are set, one
sweep is performed.
Parameters:
<NumberSweeps>
0 to 32767
*RST:
0 (GSM: 200, PHN:1)
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:POINts
<NumberPoints>
This command defines the number of measurement points to be collected during one
sweep.
Note: For Spurious Emissions measurements the maximum number of sweep points in
all ranges is limited to 100001.
Parameters:
<NumberPoints>
Range:
101 to 32001
*RST:
691
Example:
SWE:POIN 150
Manual operation:
See
See
Remote Commands in Spectrum Analyzer Mode