Remote Control
R&S
®
ESRP
795
User Manual 1175.7445.02 ─ 08
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold
<Level>
This command defines a threshold above which the signal is tracked.
Note that you have to turn on signal tracking before you can use the command
(
CALCulate<n>:MARKer<m>:FUNCtion:STRack[:STATe]
).
.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Level>
Range:
-130 dBm to 30 dBm
*RST:
-120 dBm
Default unit: dBm
Example:
CALC:MARK:FUNC:STR:THR -50DBM
Sets the threshold for signal tracking to -50 dBm.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe
<TraceNumber>
This command defines the trace on which the signal is tracked.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<TraceNumber>
Range:
1 to 6
*RST:
1
Example:
CALC:MARK:FUNC:STR:TRAC 3
Defines trace 3 as the trace for signal tracking.
Manual operation:
See
11.5.2.2
Automatic Measurement Configuration
The following commands initiate and perform automatic test routines to configure the
current measurement.
[SENSe:]ADJust:CONFiguration:HYSTeresis:LOWer
......................................................... 796
[SENSe:]ADJust:CONFiguration:HYSTeresis:UPPer
..........................................................796
[SENSe:]ADJust:CONFigure:LEVel:DURation
...................................................................796
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
........................................................ 797
...................................................................................797
..........................................................................................797
Remote Commands in Spectrum Analyzer Mode