Hiding the status bar
You can hide the status bar display, e.g. in order to enlarge the display area for the
measurement results.
1. Press the [DISPLAY] key.
2. In the "Display Settings" dialog box, select "Status Bar State: Off".
The status bar is no longer displayed.
●
To display the status bar again, select "Status Bar State: On".
SCPI command:
DISP:SBAR:STAT OFF
The following information is displayed:
Instrument status
The instrument is configured for operation with an external reference.
Error information
If errors or irregularities are detected, a keyword and an error message, if available,
are displayed in the status bar.
The following keywords are used:
UNCAL
One of the following conditions applies:
●
Correction data has been switched off.
●
No correction values are available. This occurs, for example, if a firmware
update is performed followed by a cold start of the instrument.
●
Record the correction data by performing a self alignment
OVLD
Overload of the input mixer.
●
Increase the RF attenuation (for RF input).
●
Reduce the input level (for digital input)
IFOVL
Overload of the IF signal path after the input mixer.
●
Increase the reference level.
LOUNL
Error in the instrument's frequency processing hardware was detected.
NO REF
Instrument was set to an external reference but no signal was detected on the refer-
ence input.
OVEN
OCXO reference frequency (option R&S
FSV-B4) has not yet reached its operating
temperature. The message usually disappears a few minutes after power has beeen
switched on.
Progress
The progress of the current operation is displayed in the status bar.