Remote Control
R&S
®
ESRP
683
User Manual 1175.7445.02 ─ 08
CALCulate<n>:LIMit<k>:UPPer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
If the measured values exceed the UPPer limit line, the limit is violated.
Suffix:
<n>
.
irrelevant
<k>
1...8
Selects the limit line.
Parameters:
<LimitLinePoints>
Variable number of x-axis values.
Note that the number of horizontal values has to be the same as
the number of vertical values set with
ESRP either
adds missing values or ignores surplus values.
The unit depends on
*RST:
(LIMit:STATe is set to OFF)
Example:
CALC:LIM2:UPP -10,0,0,-10,-5
Defines 5 upper limit values for limit line 2 in the preset unit.
CALC:LIM2:UPP?
Outputs the upper limit values for limit line 2 separated by a
comma.
Manual operation:
See
CALCulate<n>:LIMit<k>:UPPer:MARGin
<Margin>
This command defines an area around a upper limit line where limit check violations
are still tolerated.
Suffix:
<n>
.
irrelevant
<k>
1...8
Selects the limit line.
Parameters:
<Margin>
Numeric value that defines the margin.
In spectrum mode, the unit is always dB.
*RST:
0
Example:
CALC:LIM2:UPP:MARG 10dB
Defines the margin of limit line 2 to 10 dB below the limit value.
CALCulate<n>:LIMit<k>:UPPer:MODE
<Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
irrelevant
Remote Commands in Receiver Mode