Quick Start
R&S
®
ESRP
63
User Manual 1175.7445.02 ─ 08
Ref Level
Reference level
Att
RF attenuation level.
Ref Offset
Reference level offset
SWT
Sweep time that has been set.
If the sweep time does not correspond to the value for automatic coupling,
a bullet is displayed in front of the field. The color of the bullet turns red if
the sweep time is set below the value for automatic coupling. In addition,
the UNCAL flag is shown. In this case, the sweep time must be increased.
RBW
Resolution bandwidth that has been set.
If the bandwidth does not correspond to the value for automatic coupling,
a green bullet appears in front of the field.
VBW
Video bandwidth that has been set.
If the bandwidth does not correspond to the value for automatic coupling,
a green bullet is displayed in front of the field.
Compatible
Compatible device mode (FSP, FSU, default; default not displayed)
Mode
Indicates which sweep mode type is selected:
●
"Auto FFT": automatically selected FFT sweep mode
●
"Auto sweep": automatically selected swept sweep mode
●
"FFT": manually selected FFT sweep mode
●
"Sweep": manually selected swept sweep mode
3.4.1.3
Measurement Settings Information
In addition to the common hardware settings, the channel bar above the diagram also
displays information on instrument settings that affect the measurement results even
though this is not immediately apparent from the display of the measured values. This
information is displayed in gray font and only when applicable for the current measure-
ment, as opposed to the common hardware settings that are always displayed.
In Receiver mode, the R&S
ESRP shows the following information.
TRG
Trigger source
(for details see trigger settings in the "TRIG" menu in the Operating manual)
●
EXT
: External
●
VID
: Video
75
Ω
The input impedance of the instrument is set to 75
Ω
.
In Spectrum mode, the following types of information may be displayed, if applicable.
Basic Operations