Spectrum Measurements
R&S
®
ESRP
256
User Manual 1175.7445.02 ─ 08
Child Node
Attribute
Value
Parameter Description
Mand.
Unit
dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the stop
value
LimitFailMode
Absolute | Relative
| Absolute and Rel-
ative | Absolute or
Relative
If used, it has to be identical
to DefaultLimitFailMode
No
RBW
Bandwidth <bandwidth in Hz>
Yes
Type
NORM | PULS |
CFIL | RRC
No
VBW
Bandwidth <bandwidth in Hz>
Yes
Detector
NEG | POS |
SAMP | RMS |
AVER | QUAS
If used, it has to be identical
in all ranges.
No
Sweep
Mode
Manual | Auto
Yes
Time
<time in sec>
No
Amplitude
No
ReferenceLevel
Value
<power in dBm>
Yes, if the Refer-
enceLevel child
node is used
Unit
dBm
Defines dBm as unit
Yes, if the Refer-
enceLevel node
is used
RFAttenuation
Mode
Manual | Auto
on page 236 Yes, if the Refer-
enceLevel child
node is used
Preamplifier
ON | OFF
Yes
ASCII File Export Format (Spectrum Emission Mask)
The first part of the file lists information about the signal analyzer and the general
setup. For a detailed description refer to
Chapter 6.3.1.6, "ASCII File Export Format"
Measurements