Command Reference
R&S
®
CMW500
259
User Manual 1173.9463.02 ─ 02
Example:
SYSTem:PRESet:ALL
Force the entire R&S CMW500 to a preset state optimized for
manual operation.
*RST
Optimize the current sub-instrument for remote operation.
Usage:
Event
SCPI confirmed
Firmware/Software:
V1.0.5.3
SYSTem:TIME
<Hour>, <Minute>, <Second>
Sets the time of the internal clock.
Parameters:
<Hour>
Hour, 0 to 23
<Minute>
Minute, 0 to 59
<Second>
Second, 0 to 59 (60 due to rounding)
Numbers entered are rounded to the closest integer value.
*RST:
a *RST does not affect the time settings
Example:
SYSTem:TIME?
Query the current time. Possible response: 15, 09, 20
Usage:
SCPI confirmed
Firmware/Software:
V1.0.0.4
SYSTem:UPDate:DGRoup
<Devicegroup>
Sets the "Device Group" the instrument belongs to. For remote installation, this setting
must match the corresponding setting in the R&S Software Distributor options.
Parameters:
<Devicegroup>
Device group as string.
Example:
SYSTem:UPDate:DGRoup "MyInstrument"
Set the device group to "MyInstrument".
Firmware/Software:
V1.0.0.4
SYSTem:VERSion?
Queries the SCPI version number to which the instrument complies. The instrument
complies to the final SCPI version 1999.0.
Usage:
Query only
SCPI confirmed
Firmware/Software:
V1.0.0.4
Instrument-Control Commands