Command Reference
R&S
®
CMW-KM750/-KM751
92
User Manual 1173.9670.02 ─ 05
<EnableEMask>
OFF | ON
Disable or enable emission mask measurement
*RST:
ON
<EnableOBW>
OFF | ON
Disable or enable measurement of occupied bandwidth
*RST:
ON
Example:
See
Firmware/Software:
V3.0.10
V3.0.30: maximum number of segments increased to 500
Options:
R&S CMW-KM012
TRIGger:TDSCdma:MEAS<i>:MEValuation:LIST:MODE
<TriggerMode>
Specifies the trigger mode for list mode measurements. For configuration of retrigger
flags see
Parameters:
<TriggerMode>
ONCE | SEGMent
ONCE:
A trigger event is only required to start the measurement.
As a result the entire range of segments to be measured is cap-
tured without additional trigger event. The retrigger flags of the
segments are ignored.
SEGMent:
The retrigger flag of each segment is evaluated. It
defines whether the measurement waits for a trigger event before
capturing the segment, or not.
*RST:
ONCE
Example:
See
Firmware/Software:
V3.0.10
Options:
R&S CMW-KM012
6.3.5 Trigger Settings
The following commands define the trigger parameters.
...........................................93
..........................................................93
.............................................................93
.......................................................94
...........................................................94
...........................................................94
..............................................................94
..............................................................95
Multi Evaluation Measurement Commands