Command Reference
R&S
®
CMW-KM645
53
User Manual 1173.9740.02 ─ 04
Return values:
<Reliability>
See
<OutOfTol>
Out of tolerance result, i.e. percentage of measurement intervals
in the statistics cycle (
.
Range:
0 % to 100 %
Default unit: %
<CarrierPower>
RF carrier power
Range:
-130 dBm to +55 dBm
Example:
See
Performing Single Shot RF Measurements
Usage:
Query only
Firmware/Software:
V1.0.15.20
FETCh:FMSTereo:MEAS<i>:MEValuation:RFCarrier:SDEViation?
READ:FMSTereo:MEAS<i>:MEValuation:RFCarrier:SDEViation?
CALCulate:FMSTereo:MEAS<i>:MEValuation:RFCarrier:SDEViation?
Returns the standard deviation of the "RF Carrier" results.
The values described below are returned by
FETCh
and
READ
commands.
CALCulate
commands return limit check results instead, one value for each result listed below.
The number to the left of each result parameter is provided for easy identification of the
parameter position within the result array.
Return values:
<1_Reliability>
See
<2_OutOfTol>
Out of tolerance result, i.e. percentage of measurement intervals
in the statistics cycle (
.
Range:
0 % to 100 %
Default unit: %
<3_CarrierPower>
Standard deviation of the carrier power
Range:
0.0 dB to 92.5 dB
Default unit: dB
<4_CarrierFreqErr>
Standard deviation of the carrier frequency error
Range:
0 Hz to +96 kHz
Default unit: Hz
<5_MultDevPosNeg> Standard deviation of the multiplex deviation,
±
Peak / 2
Range:
0 Hz to +48 kHz
Default unit: Hz
Measurement Results
深圳德标仪器
135-1095-0799