Remote Control Commands
R&S
®
AREG100A
244
User Manual 1178.7417.02 ─ 04
*RST:
n.a. (factory preset: 0)
[:SOURce]:FREQuency[:CW|FIXed]?
<Frequency>
Gets the frequency of the RF output signal in the selected path.
In CW mode the instrument operates at a fixed frequency (24 MHz or 77 MHz).
Parameters:
<Frequency>
float
Usage:
Query only
[:SOURce<hw>]:FREQuency[:CW|FIXed]:RCL
<Rcl>
Set whether the RF frequency value is retained or taken from a loaded instrument con-
figuration, when you recall instrument settings with command
*RCL
.
Parameters:
<Rcl>
INCLude | EXCLude
INCLude
Takes the frequency value of the loaded settings.
EXCLude
Retains the current frequency when an instrument configuration
is loaded.
*RST:
INCLude
Example:
SOURce1:FREQuency:CW:RCL INCLude
Manual operation:
See
[:SOURce<hw>]:AREGenerator:OBJect:ALL[:STATe]
<GlobalObjStat>
Switches all available radar objects (1 to 4) on or off simultaneously.
Parameters:
<GlobalObjStat>
0 | 1 | OFF | ON
Switches all available radar objects on or off.
*RST:
0
Example:
:SOURce:AREGenerator:OBJect:ALL:STATe 1
Manual operation:
See
[:SOURce<hw>]:AREGenerator:OBJect<ch>:ATTenuation
<AregObjAtt>
Defines the attenuation of a specific radar object. Together with the base attenuation
valid for all radar objects, it forms the total attenuation for the specific object.
Suffix:
<ch>
.
1 to 4
Defines the specific radar object.
SOURce Subsystem