Remote Control Commands
R&S
®
AREG100A
237
User Manual 1178.7417.02 ─ 04
Example:
See
Usage:
Query only
:SYSTem:SPECification:VERSion:FACTory?
Queries the data sheet version of the factory setting.
Return values:
<Version>
string
Example:
See
Usage:
Query only
Manual operation:
See
:SYSTem:SRData?
Queris the SCPI recording data from the internal file.
This feature enables you to transfer an instrument configuration to other test environ-
ments, as e.g. laboratory virtual instruments.
Return values:
<FileData>
block data
Example:
SYSTem:SRData?
// #3118:SOURce1:ROSCillator:SOURce EXT
// :SOURce1:FREQuency:CW 4000000000
// :SOURce1:FREQuency:OFFSet 1000000
// :SOURce1:AM1:STATe 1
// :OUTPut1:STATe 1
Usage:
Query only
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:STAR:COMP?
Response:
1
the startup of the instrument is completed.
Usage:
Query only
:SYSTem:DATE
<Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
SYSTem Subsystem