Getting Started
R&S
®
AREG100A
19
User Manual 1178.7417.02 ─ 04
Mounting in a rack
The R&S
AREG100A can be installed in a rack using a rack adapter kit (Order No. see
data sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation concept to
ensure that the instruments do not overheat. Insufficient airflow for a longer period can
disturb the operation and even cause damage.
3.1.1.5
Connecting AC Power
The R&S
AREG100A is equipped with an AC power supply connector, that can be
operated with different AC power voltages. Once it is connected, the instrument auto-
matically adjusts to the given voltage. Refer to the data sheet for the requirements of
voltage and frequency. There is no need to set the voltage manually or change fuses.
To connect the AC supply
► Connect the R&S
AREG100A to the AC power source using the supplied power
cable.
Note:
Since the instrument is designed in compliance with standard EN 61010-1
safety class I, it must only be connected to an outlet that has a ground contact.
Characteristics of the AC power supply:
●
100 V to 240 V AC
3.1.1.6
Cabling of Front End and Base Unit
To connect the front end
1. Connect the IF cable with the label RX IF OUT to the Rx IF Out connector of the
front end.
2. Connect the IF cable with the label TX IF IN to the Tx IF IN connector of the front
end.
3. Connect the control cable to the Control connector of the front end.
Preparing for Use