Instrument tour
R&S
®
RTP
39
Getting Started 1803.6478.02 ─ 01
Press the [Intensity] key and turn the [Multiuse] knob. Press the [Intensity] key
again to toggle between the three settings. The controlled parameter and its value
are shown in the input box in the upper right corner of the screen.
[Cursor]
Starts a cursor measurement: displays vertical and horizontal cursor lines in the
active diagram, and the results of the cursor measurement.
If you press the key while a cursor measurement is enabled, the "Cursors" dialog
opens. The next press closes the dialog.
Cursors are markers which are placed at points of interest on a waveform. The
instrument measures the cursor positions and delta values between parallel cur-
sors.
[Measure]
Opens and closes the "Measurements" dialog and the subtab of the last used
measurement. In the dialog, you can configure and start automatic measure-
ments.
[History]
Activates the history and opens the history player. The second press opens the
player configuration. The history shows stored acquisitions that were acquired
before the current one.
The key is illuminated as long as the history is active.
[HD]
Activates the high definition mode and opens the "HD Mode" dialog box.
[Touch Lock]
Locks the touchscreen to prevent unintended use. When the touchscreen is off,
the key is illuminated. Press again to unlock the touchscreen.
Camera
Performs the action that is assigned to the key in "Menu" > "Settings" > "Frontpa-
nel" > "Hardkeys". By default, the key saves a screenshot of the waveform dis-
play.
Keys and controls