Instrument tour
R&S
®
RTP
34
Getting Started 1803.6478.02 ─ 01
[Auto Norm]
Toggles the trigger mode between "Auto" and "Normal". The current setting is
shown on the trigger label.
Auto
The instrument triggers repeatedly after a time interval if the
trigger conditions are not fulfilled.
Norm
The instrument acquires a waveform only if a trigger occurs.
The "Norm" LED lights up in green.
[Source]
Opens a dialog box where you can select the trigger source. Press the key again
to switch the source. The key lights up in the color of the selected trigger source.
[Slope]
Toggles the trigger slope or trigger polarity, dependent on the trigger type. The
current setting is shown on the trigger label.
[Zone]
Opens the overlay menu to configure the zone trigger. A zone trigger combines
the trigger condition with the intersection or non-intersection of one or more zones
or masks.
5.3.3
Horizontal controls
The rotary knobs in the Horizontal functional block adjust the horizontal parame-
ters. These settings are effective for all channel waveforms. In addition, a [Zoom]
is available.
[Scale]
The rotary knob adjusts the time scale for all signals. The time scale is also
known as timebase.
Turn clockwise to stretch the waveforms. Doing so, the scale value
time/div
decreases.
Press the knob to toggle between coarse and fine scale adjustment.