Remote control basics
R&S
®
NRX
488
User Manual 1178.5566.02 ─ 09
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
0
Sensor A calibrating
Sensor B calibrating
Sensor C calibrating
Sensor D calibrating
NRT Sensor calibrating
Other Sensor calibrating
0
0
0
0
0
0
0
0
0
+
Figure 14-10: Operation calibrating status register
Querying the register:
●
STATus:OPERation:CALibrating[:SUMMary]:CONDition?
●
STATus:OPERation:CALibrating[:SUMMary][:EVENt]?
Table 14-12: Used operation calibrating status bits and their meaning
Bit no.
Short description
Bit is set if a sensor is zeroing.
1
Sensor
A calibrating
Sensor
A
2
Sensor
B calibrating
Sensor
B
3
Sensor
C calibrating
Sensor
C
4
Sensor
D calibrating
Sensor
D
5
NRT sensor calibrating
R&S
NRT
‑
Zxx power sensor
6
Other sensor calibrating
USB or LAN power sensor
14.4.8.2
Operation measuring status register
The
CONDition
register contains information whether a power sensor is measuring.
The
EVENt
register contains information whether a measurement was started or com-
pleted since the last query.