Remote Control Commands
R&S
®
NRX
313
User Manual 1178.5566.02 ─ 08
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
CALCulate<Measurement>:STATistics:MARKer:VERTical:DATA[:POWer]?
Effective for statistics measurements.
Queries the power value at the y-marker position.
The command is not fully R&S NRP2 compatible. If port
1 is assigned to measure-
ment
1 and so on, the command works. If port
1 is assigned to measurement
2, for
example, enable the R&S NRP2 emulation using
"NRP2"
.
Suffix:
<Measurement>
.
1 to 8
Measurement
Usage:
Query only
CALCulate<Measurement>:STATistics:MARKer:X:POSition[:ABSolute]
<value>
Effective for statistics measurements.
Sets the absolute position of the x-marker.
Suffix:
<Measurement>
.
1 to 8
Measurement
Parameters:
<value>
Range:
-200.0 to 200.0
*RST:
0.0
Default unit: dBm
Manual operation:
See
CALCulate<Measurement>:STATistics:MARKer:X:POSition:RELative
<value>
Effective for statistics measurements.
Sets the x-marker relative to a power value.
Suffix:
<Measurement>
.
1 to 8
Measurement
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH