Preparing for Use
R&S
®
QAR
30
User Manual 1178.7200.02 ─ 09
b) If normalization still indicates a misalignment, repeat the adjustment until the
error no longer occurs.
4.2.5
Installing the R&S
QAR on the Platform (R&S
QAR-Z21)
Required tools:
●
Allen wrench size 5 (for screws ISO 10642 M8x20)
●
Allen wrench size 6 (for screws ISO 4762 M8x40)
●
Allen wrench size 10 (for screws ISO 4762 M12x55)
●
Open-end wrench size 19
●
Spirit level.
Installing the platform
1. Place the platform on a stable and horizontal surface that can support the weight of
the R&S
QAR.
2. Check with a spirit level if the platform itself is level.
3. If necessary, remove any irregularities by adjusting the height of the feet until the
platform is level.
● A clockwise movement lowers the corresponding platform area.
● A counterclockwise movement raises the corresponding platform area.
4. Tighten the feet with the check nuts using the open-end wrench.
Figure 4-12: Platform alignment
Mounting the R&S
QAR
When the platform is in place, you can mount the R&S
QAR on the platform.
Installing the R&S
QAR