Remote control commands
R&S
®
RTM3000
574
User Manual 1335.9090.02 ─ 09
Use this command only for envelope waveforms. For other channel waveforms use
.
.
To set the range of samples to be returned, use
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<EnvelopeData>
List of values according to the format settings - the voltages of
the envelope points. The list contains two values for each sam-
ple interval.
Usage:
Query only
CHANnel<m>:DATA:ENVelope:HEADer?
Returns information on the envelope waveform.
Use this command only for envelope waveforms. for all other channel waveforms use
.
Table 17-2: Header data
Position
Meaning
Example
1
XStart in s
-9.477E-008 = - 94,77 ns
2
XStop in s
9.477E-008 = 94,77 ns
3
Number of samples
120000
4
Number of values per sample interval. For envelope
waveforms the value is 2.
2
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<EnvelopeHeader>
Comma-separated value list
Example:
-9.477E-008,9.477E-008,200000,2
Usage:
Query only
17.9.1.3
Math waveforms
In addition to the commands described below, consider also the following commands:
●
CALCulate:MATH<m>:DATA:XINCrement?
●
CALCulate:MATH<m>:DATA:XORigin?
Documenting results