Instrument Tour

R&S

®

RTC1000

14

Getting Started 1335.7346.02 ─ 04

4.1.1

 

Input Connectors

BNC inputs (6, 7)

The R&S

 

RTC1000 has two channel inputs (6) to connect the input signals. The exter-

nal trigger input (7) is used to control the measurement by an external signal. The trig-
ger level can be set from -5

 

V to 5

 

V.

The input impedance of all BNC inputs is 1

 

M

Ω

.

Risk of electrical shock - maximum input voltages

The maximum input voltage on 

channel inputs

 must not exceed 200

 

V

 

(peak) and

150

 

V

 

(RMS).

For the 

external trigger input

, the maximum input voltage is 100

 

V

 

(peak) and

70

 

V

 

(RMS).

Transient overvoltages must not exceed 200

 

V

 

(peak).

Voltages higher than 30

 

V

 

(RMS) or 42

 

V

 

(peak) or 60

 

V

 

DC are regarded as hazard-

ous contact voltages. When working with hazardous contact voltages, use appropriate
protective measures to preclude direct contact with the measurement setup:

Use only insulated voltage probes, test leads and adapters.

Do not touch voltages higher than 30

 

V

 

(RMS) or 42

 

V

 

(peak) or 60

 

V

 

DC.

Risk of injury and instrument damage

The instrument is not rated for any measurement category. When measuring in circuits
with transient overvoltages of category II, III or IV circuits, make sure that no such
overvoltages reach the R&S

 

RTC1000 input. Therefore, use only probes that comply

with DIN EN 61010-031. When measuring in category II, III or IV circuits, always insert
a probe that appropriately reduces the voltage so that no transient overvoltages higher
than 200

 

V

 

(peak) are applied to the instrument. For detailed information, refer to the

documentation and safety information of the probe manufacturer.
Explanation: According to section AA.2.4 of EN 61010-2-030, measuring circuits with-
out any measurement category are intended for measurements on circuits which are
not directly connected to the mains.

Front Panel

Содержание 1335.7500K02

Страница 1: ...R S RTC1000 Digital Oscilloscope Getting Started Getting Started Version 04 1335734602 S 2...

Страница 2: ...4 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Data without tolerance limits is not binding R S is a registered trademark of Rohde Schwarz GmbH Co KG Trade names are t...

Страница 3: ...4 Release Notes Open Source Acknowledgment 8 3 Preparing for Use 9 3 1 Unpacking and Checking the Instrument 9 3 2 Positioning the Instrument 9 3 3 Starting the Instrument 10 3 3 1 Powering On 10 3 3...

Страница 4: ...Contents R S RTC1000 4 Getting Started 1335 7346 02 04...

Страница 5: ...is part of the product documentation It warns you about the poten tial dangers and gives instructions how to prevent personal injury or damage caused by dangerous situations Safety information is prov...

Страница 6: ...wall is 10 cm Signal levels at the input connectors are all within the specified ranges Signal outputs are correctly connected and are not overloaded Instrument damage caused by electrostatic discharg...

Страница 7: ...etting started manual The online version of the user manual provides the complete contents for immediate display on the internet Instrument help The help offers quick context sensitive access to the f...

Страница 8: ...alibration Certificate The document is available on https gloris rohde schwarz com calcert You need the device ID of your instrument which you can find on a label on the rear panel 2 4 Release Notes O...

Страница 9: ...u can use the material to protect the control elements and connectors Delivery contents The delivery package contains the following items R S RTC1000 digital oscilloscope R S RT ZP03 probe probes 2x C...

Страница 10: ...e R S RTC1000 to overheat which can impair the measurement results disturb the operation and even cause damage Ensure that all fan openings are unobstructed and that the airflow perforations are unimp...

Страница 11: ...Shutting Down To start up the instrument 1 Make sure that the R S RTC1000 is connected to the AC power supply and the main power switch on the rear panel is in position I 2 Press the ON OFF key The k...

Страница 12: ...can be lost Press the ON OFF key first to shut down the application properly 3 4 Replacing the Fuse The instrument is protected by a fuse You can find it on the rear panel between the main power swit...

Страница 13: ...ront view of the R S RTC1000 1 Display 2 Softkeys and menu 3 ON OFF key 4 Cursor Menu Analyze and General sections 5 Vertical Trigger and Horizontal sections 6 BNC connectors of the analog channel inp...

Страница 14: ...to preclude direct contact with the measurement setup Use only insulated voltage probes test leads and adapters Do not touch voltages higher than 30 V RMS or 42 V peak or 60 V DC Risk of injury and in...

Страница 15: ...RTC B1 Connecting other probe types can demolish the input 4 1 2 Other Connectors on the Front Panel Pattern Generator 8 Connectors for the pattern generator P0 P1 P2 P3 Probe Comp 8 Probe compensati...

Страница 16: ...1 The instrument supports a wide range power supply It automatically adjusts to the cor rect range for the applied voltage There is no line voltage selector The AC main power switch disconnects the i...

Отзывы: