Command Reference
R&S
®
ZNA
890
User Manual 1178.6462.02 ─ 12
SLOPe - return the slope (difference) between two marker val-
ues.
FLATness - return the flatness of the trace between two marker
positions.
ALL - return all statistical values, observing the order used
above.
The data is returned as a comma-separated list of real numbers.
The unit is the default unit of the measured parameter (see
CALCulate<Ch>:PARameter:SDEFine
). For com-
plex traces the statistical results MEAN, STDDev, MAX, MIN,
RMS and PTPeak are calculated in the selected format (see
CALCulate<Chn>:STATistics:FORMat
Example:
*RST; :CALC:STAT:RES? MAX
Calculate and return the maximum of the default trace showing
an S-parameter on a dB Mag scale.
:CALC:FORM POL; STAT:RES? MAX
Display the trace in a polar diagram and re-calculate the maxi-
mum. The result corresponds to the previous result but is con-
verted to a unitless linear value.
Usage:
Query only
Manual operation:
See
"Min/Max/Peak-Peak, Mean/Std Dev/RMS"
CALCulate<Chn>:STATistics:RMS[:STATe]
<Boolean>
CALCulate<Chn>:STATistics:SFLatness[:STATe]
<Boolean>
These commands display or hide the "RMS" and the "Flatness/Gain/Slope" results in
the diagram area of trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON | OFF - statistical info field on or off.
*RST:
OFF
Example:
See
CALCulate<Chn>:STATistics[:STATe]
Manual operation:
See
8.3.1.21
CALCulate:TDIF...
Commands related to
.
CALCulate<Ch>:TDIF:IMBalance:COMPensation[:STATe]
<Boolean>
Selects the calculation method for S-parameters, ratios and derived quantities during
an amplitude imbalance or phase imbalance sweep.
SCPI Command Reference