Command Reference
R&S
®
ZNA
873
User Manual 1178.6462.02 ─ 12
CALCulate<Chn>:RIPPle:CONTrol:DOMain
<SweepType>
Deletes the existing ripple limit ranges and (re-)defines the physical units of the stimu-
lus values of the ripple limit lines. The unit of the ripple limit is defined via
CALCulate<Chn>:RIPPle:RDOMain:FORMat
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Setting parameters:
<SweepType>
FLIN | FLOG | FSEG | FSINgle | TLIN | TLOG | PLIN | PLOG |
PSINgle
Keywords for the units of the stimulus values; frequency, power,
and time units.
The selected unit must be compatible with the sweep type (see
FLOG, FSEG and FSINgle, s for TLIN and TLOG, dBm for PLIN,
PLOG and PSINgle. Otherwise the ripple limit lines cannot be
displayed and no ripple limit check is possible.
*RST:
FLIN
Example:
SWE:TYPE POW
Select a power sweep.
CALC:RIPP:CONT:DOM PLIN
Delete all existing ripple limit ranges and select level units for the
domain of the active trace.
CALC:RIPP:DATA 1, -10, -5, 3
Define and enable a ripple limit range in the stimulus range
between -10 dBm and -5 dBm, assigning a ripple limit of 3 dB.
Usage:
Setting only
Manual operation:
See
"Add / Insert / Delete / Delete All / Align All"
CALCulate<Chn>:RIPPle:DATA
<RippleLimRange>...
Adds and enables/disables an arbitrary number of ripple limit ranges, assigning the
stimulus values and the ripple limits. See
Chapter 5.4.1.2, "Rules for Ripple Test Defini-
Note:
This command does not overwrite existing ripple limit ranges. The defined
ranges are appended to the range list as new ranges. Use the
CALCulate<Chn>:RIPPle:SEGMent<Seg>...
commands to change existing ripple
limits.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<RippleLimRange>
Parameter list in the format <Type>, <StartStimulus>, <StopSti-
mulus>, <RippleLimit>[, {<Type>, <StartStimulus>, <StopStimu-
lus>, <RippleLimit>}], where:
SCPI Command Reference