GUI Reference
R&S
®
ZNA
702
User Manual 1178.6462.02 ─ 12
Type
The name of the converter type.
Remote command:
[SENSe:]CONVerter:DEFinition:CATalog?
[SENSe:]CONVerter:DEFinition:DEFine
IF
Defines the fixed IF frequency of the converter type.
Remote command:
[SENSe:]CONVerter:DEFinition:IFRequency
Converter Port: Start Freq./Stop Freq.
Defines the start and stop frequency of the (converter port of the) converter type.
Remote command:
[SENSe:]CONVerter:DEFinition:FREQuency
Source: Multiplicator
Defines the source multiplicator, i.e. the factor by which the incoming RF frequency is
multiplied by the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:SMUL
Source: Nominal Power
Defines the preferred input power at the RF In port of the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:SPOWer
Source: Converter Max. Power
Defines the maximum tolerable input power at the RF In port of the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:MPOWer
Local Port: Multiplicator
Defines the source multiplicator, i.e. the factor by which the incoming LO frequency is
multiplied by the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:LMUL
Local Port: Nominal Power
Defines the preferred input power at the LO In port of the converter.
Remote command:
[SENSe:]CONVerter:DEFinition:LPOWer
Add/Copy
Use "Add" to create a converter type with default properties or use "Copy" to duplicate
the properties of the converter type whose tab is currently selected.
Setup Softtool