Getting Started
R&S
®
ZNA
57
User Manual 1178.6462.02 ─ 12
Select Trace – [Trace
Config] and use the control elements in the "Traces" softtool tab
if you wish to create additional traces and diagrams.
4.4.1.2
Selecting the Sweep Range and Other Parameters
After a system preset the display shows a diagram with a dB magnitude scale, and the
S-parameter S
21
is selected as a measured quantity. This S-parameter is the forward
transmission coefficient of the DUT. It is defined as the ratio of the transmitted wave at
the DUT's output port (port no. 2) to the incident wave at the DUT's input port (port no.
1).
The R&S
ZNA automatically adjusts its internal source and receiver to the selected
measured quantities: For an S
21
measurement, a stimulus signal (termed a
1
) is trans-
mitted at the analyzer port no. 1; the transmitted wave (termed b
2
) is measured at port
2. The stimulus signal from the analyzer port no. 2 is not needed except for some cali-
bration types.
By default the sweep range is set to the frequency range of the analyzer, which can be
unsuitable for your DUT. The following procedure shows you how to configure a
smaller sweep range.
1. Select Stimulus – [Start] and set the "Start Frequency" to the lowest frequency you
want to measure (e.g. 1.77 GHz).
2. In the "Stop Frequency" input field, enter the highest frequency you want to mea-
sure (e.g. 2.5 GHz).
3. Select Trace – [Scale] > "Scale Values" and activate the "Auto Scale Trace" func-
tion. The analyzer adjusts the scale of the diagram to fit in the entire S
21
trace,
leaving an appropriate display margin.
Tip:
Refer to
Chapter 4.3.7, "Scaling Diagrams"
on page 50 to learn more about
the different methods and tools for diagram scaling.
4.4.1.3
Calibrating the Instrument
Calibration (system error correction) is the process of eliminating systematic, reprodu-
cible errors from the measurement results. E.g., in the current test setup, the connect-
ing cables between the analyzer ports and the DUT introduce an attenuation and a
Performing Measurements