Channel table
The rows and columns of the channel table represent the existing channels (rows)
together with certain editable (white) or non-editable (gray) properties (columns).
●
"Name" indicates the name of the related channel.
●
"Traces" indicates the names of all traces assigned to the related channel.
●
"On/Off" toggles the measurement state of the related channel.
●
"Single Sweep" toggles between "Continuous" and "Single" sweep mode (see
Remote command:
CONFigure:CHANnel:CATalog?
CONFigure:CHANnel<Ch>:NAME
CONFigure:CHANnel<Ch>:NAME:ID?
CONFigure:CHANnel<Ch>:MEASure[:STATe]
INITiate<Ch>:CONTinuous
Add / Delete
The buttons below the channel table add and delete channels.
●
"Add" adds a new channel to the list. The new channel is named Ch<n>, where
<n> is the largest of all existing channel numbers plus one.
●
"Delete" deletes the channel selected in the table. This button is disabled if the
setup contains only one channel: In manual control, each setup must contain at
least one diagram area with one channel and one trace.
Remote command:
6.12.2
Port Config Tab
Configures the source and receive ports of the R&S
ZNA, for measurements on fre-
quency-converting DUTs, defining arbitrary port frequencies and powers. While the
arbitrary mode is active, "Arb" and the selected stimulus axis are displayed in the chan-
nel line.
The arbitrary port configuration is part of the "Frequency Conversion Measurements"
option R&S
ZNA-K4. If this option is not equipped, the "Port Config" Tab is not availa-
ble.
Background information
Refer to
Chapter 5.7.3, "Frequency Conversion Measurements"
Channel Config Softtool