GUI Reference
R&S
®
ZNA
489
User Manual 1178.6462.02 ─ 12
"Power Meter": Use a power meter for the calibration of the reference receiver
and
the
flatness calibration.
For external generators and the optional
on page 224 port, a
power meter must be used.
Remote command:
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:METHod
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the power calibration of a source port. The last configured power meter is selected
by default. See
Remote command:
SOURce:POWer:CORRection:PMETer:ID
Auto Zero
Initiates an automatic zeroing procedure of the power meter; see
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:AZERo
6.11.3.2
Cal Power Config Dialog
The "Cal Power Config" dialog adjusts the target power for the power calibration (cal
power) and defines the target power for the reference receiver calibration. These set-
tings are particularly important for test setups involving external attenuators or amplifi-
ers.
Access:
Channel – [Cal] > "Pwr Cal Settings" > "Cal Power..."
Related information
The settings in the "Cal Power Config" dialog are also used to define port-specific
source powers in arbitrary mode; refer to
Cal Softtool