GUI Reference
R&S
®
ZNA
260
User Manual 1178.6462.02 ─ 12
To configure external generators, use the "External Generators" dialog. See
ter 5.7.28, "External Generators"
Properties
Allows you to select the measurement detector and the display unit for the selected
wave quantity.
Detector ← Properties
Selects the algorithm that is used to calculate the results points from the raw measure-
ment data.
For details refer to
Chapter 5.3.5.3, "Detector Settings"
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:SDEFine
[SENSe<Ch>:]SWEep:DETector:TIME
Display unit ← Properties
Selects the physical unit of the displayed trace. It is possible to display the measured
"Voltage" V or to convert it into a power value P according to the formula
P = V
2
/ Re(Z
0
).
Z
0
denotes the reference impedance of the source port (for wave quantities a
n
) or of
the receive port (for wave quantities b
n
). The reference impedances are defined in the
"Balanced Ports" dialog; see
Chapter 6.2.2.5, "Balanced Ports Dialog"
Remote command:
6.2.4
Ratio Tab
Selects ratios of wave quantities as measured quantities.
Background information
Refer to
Chapter 5.3.5, "Wave Quantities and Ratios"
Meas Softtool