GUI Reference
R&S
®
ZNA
246
User Manual 1178.6462.02 ─ 12
S<out><in>
Selects one of the four elements of the standard 2-port
quantity for the active trace.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22"
All S-Params
With the active logical ports distributed to m DUTs with n
1
, ..., n
m
ports, respectively, the
analyzer creates n
1
2
+ ... + n
m
2
diagrams and displays the full set of
, one
in each diagram. The diagrams are arranged as an n x n grid.
If "All S-Params" would result in more than 100 diagrams being created, a warning dia-
log is opened that lets you cancel the diagram creation.
The reflection coefficients S
ii
appear in the diagrams on the main diagonal, the trans-
mission coefficients S
ij
(i
≠
j) occupy the other diagrams. By default, reflection coeffi-
cients are displayed in Smith diagrams; transmission coefficients in Cartesian dia-
grams "dB Mag" scale.
See also
.
Remote command:
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:DEFine:SGRoup
More S-Params...
Opens a dialog that allows to select an arbitrary S-parameter as measured quantity,
along with a suitable
Chapter 6.2.2.3, "More S-Params Dialog"
Y/Z/k/μ-Parameters
Allows you to select other measurement results that are typically used to characterize
a linear DUT. All these results can be calculated from the DUT's S-parameters.
Z <- S-Parameters ← Y/Z/k/μ-Parameters
Allows you to select a converted impedance as measured quantity.
A converted impedance is calculated from the corresponding S-parameter, which can
be selected in the pull-down lists below "Z ←". In case a transmission parameter is
selected, you can choose between a
impedance calculation.
Meas Softtool