GUI Reference
R&S
®
ZNA
245
User Manual 1178.6462.02 ─ 12
Setup
.
S-Parameter (selector)
as a measured quantity for the active trace.
Single-ended (unbalanced) S-parameters are referred to as S<out>< in>, where <out>
and <in> denote the output and input
logical
port numbers, respectively.
In presence of balanced ports, standard S-parameters are defined in the form
S<m_out><m_in><out><in>, where output mode <m_out> and input mode <m_in> can
be one of:
●
d (differential, balanced)
●
c (common, balanced)
●
s (single-ended, unbalanced)
It is also possible to display "raw" single-ended S-parameters within the same channel:
after setting the S-Parameter type to "S" any pair of (used)
physical
ports can be
selected.
Remote command:
CALCulate<Ch>:PARameter:MEASure
CALCulate<Ch>:PARameter:MEASure:SENDed
CALCulate<Ch>:PARameter:SDEFine
CALCulate<Ch>:PARameter:SDEFine:SENDed
Meas Softtool