
Command Reference
R&S
®
ZNA
1363
User Manual 1178.6462.02 ─ 12
Parameters:
<FixedFrequency>
Fixed frequency.
Range:
Depending on the instrument model.
Increment: 0.1 kHz
*RST:
Minimum of the analyzer's frequency range, fMIN
Default unit: Hz
Example:
*RST; FREQ:CONV:MIX:RFF 1 GHz
Reset the analyzer and specify a fixed frequency of 1 GHz, to be
assigned to the RF signal.
FREQ:CONV:MIX:FFIX?; FIX?
Query the fixed frequency and the signal assignment using the
alternative commands. The response is 1000000000;RF.
[SENSe<Ch>:]FREQuency:IMODulation:TTOutput
<TwoToneOutput>
Selects the source for the two tone output signal for intermodulation measurements.
Deprecated, superseded by
Suffix:
<Ch>
.
Channel number.
Parameters:
<TwoToneOutput>
PORT | EDEVice
PORT
Analyzer port 1.
Requires a 4-port R&S
ZNA with direct generator/receiver option
R&S
ZNAxx-B16.
EDEVice
External combiner
*RST:
EDEVice
Example:
See
[SENSe<Ch>:]FREQuency:IMODulation:LTONe
[SENSe<Ch>:]FREQuency:MODE
<FreqSweep>
Selects the sweep type and defines which set of commands controls the stimulus fre-
quency.
Tip:
The command
provides a complete list of sweep
types.
Suffix:
<Ch>
.
Channel number
R&S ZVR/ZVABT Compatible Commands