Command Reference
R&S
®
ZNA
1354
User Manual 1178.6462.02 ─ 12
Parameters:
<Channel>
CHANnel1 | CHANnel2 | CHANnel3 | CHANnel4
Number of the channel to be activated. The channel must be
created before using
ON
.
*RST:
CHANNEL1
Example:
CONF:CHAN2:STAT ON; :INST CHANnel2
Create channel no. 2 and select it as the active channel.
OUTPut<Chn>:DPORt
<Port>
Selects a source port for the stimulus signal (drive port). The setting acts on the active
trace. The effect of the drive port selection depends on the measurement parameter
associated to the active trace:
●
If an S-parameter S
<out><in>
is measured, the second port number index <in> (input
port of the DUT = drive port of the analyzer) is set equal to the selected drive port:
Drive port selection affects the measured quantity.
●
If a wave quantity or a ratio is measured, the drive port is independent from the
measured quantity:
Note:
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Port>
PORT<no>
Physical port number.
*RST:
PORT1
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'A1'
Create channel 4 and a trace named "Ch4Tr1" to measure the
wave quantity a
1
. The trace automatically becomes the active
trace.
OUTP4:DPOR PORT2
Select drive port 2 for the active trace.
[SENSe<Ch>:]CORRection:CKIT:<ConnType>:<StandardType>
<CalkitName>,
<Standard>, <MinFreq>, <MaxFreq>, <DelayParam>, <Loss>, <C0>, <L0>,
<C1>, <L1>, <C2>, <L2>, <C3>, <L3>, OPEN | SHORt
Defines the parameters of a calibration standard
<StandardType>
for a specified
connector type
<ConnType>
. A particular physical standard can be selected by speci-
fying the name of the calibration kit and its serial number. Depending on the standard
type, only a subset of the parameters may be used; see
.
Note:
If the specified cal kit does not exist, it is created with the specified calibration
standard.
R&S ZVR/ZVABT Compatible Commands