SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPower
<Power>
Defines the external generator power value after the end of a sweep. The setting is rel-
evant if the generator is switched to a user-selected power; see example.
Suffix:
<Gen>
.
Generator number
Parameters:
<Power>
End power
Range:
-145 dBm to +30 dBm
*RST:
0 dBm
Default unit: dBm
Example:
SYSTem:COMMunicate:RDEVice:GENerator:SEPMode
USER
Switch the generator to a user-selected power value after the
sweep is terminated.
SYSTem:COMMunicate:RDEVice:GENerator:SEPower
-10
Define a end power of -10 dBm.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer:CATalog?
Queries the numbers of all configured external power meters. The response is a string
containing a comma-separated list of power meter numbers.
Example:
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter 1',
'NRP-Z55', 'usb', '100045'
SYST:COMM:RDEV:PMET2:DEF 'USB Power Meter 2',
'NRP-Z55', 'usb', '100046'
Configure two R&S NRP power meters as external power meter
no. 1 and 2, assigning names and serial numbers.
SYST:COMM:RDEV:PMET:CAT?
Query the power meter numbers. The response is '1, 2'.
SYST:COMM:RDEV:PMET:COUN?
Query the number of configured power meters. The response is
2.
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer:CONFigure:AUTO[:STATe]
<Boolean>
Enables or disables auto-configuration of R&S
NRPxxS/SN power sensors. If the func-
tion is enabled, the analyzer automatically configures all power sensors detected at
any of the USB ports as Pmtr 1, Pmtr 2, ...
Parameters:
<Boolean>
ON | OFF – enable or disable auto-configuration.
SCPI Command Reference