Command Reference
R&S
®
ZNA
1284
User Manual 1178.6462.02 ─ 12
<CalkitIndex>
For calibration units that are equipped with more than one cali-
bration kit, this parameter indicates the 1-based index of the cal
kit to be used (mandatory!).
For other calibration units, this parameter is optional, but if it is
provided, it must be set to 1.
Example:
SYST:COMM:AKAL:CONN THR, 1, 2
Connect a through standard between ports 1 and 2 of the cal
unit.
Usage:
Setting only
SYSTem:COMMunicate:AKAL:MMEMory[:STATe]
<Boolean>
Shows or hides the memory of the active calibration unit (see
COMMunicate:RDEVice:AKAL:ADDRess
).
Setting parameters:
<Boolean>
ON - memory is shown in a separate drive.
OFF - memory is not shown.
*RST:
OFF
Example:
SYST:COMM:AKAL:MMEM ON
Show the memory of the active calibration unit.
Usage:
Setting only
SYSTem:COMMunicate:CODec
<Codec>
Selects the character encoding used at the remote interface. The selected encoding
applies to directory and file names, CalKit names, CalUnit characterizations and dis-
play titles.
Parameters:
<Codec>
ASCii | UTF8 | SJIS
ASCii
: 8-bit ANSI (default)
UTF8
: UTF-8
SJIS
: Shift JIS
Manual operation:
See
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
<GBIBAddress>
Sets the GPIB address of the analyzer.
Parameters:
<GBIBAddress>
GPIB address (integer number)
Range:
0 to 30
SCPI Command Reference