Command Reference
R&S
®
ZNA
1263
User Manual 1178.6462.02 ─ 12
<ListNo>
Number of point in the list.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Manual operation:
See
"Frequency / Transm. Coefficients: Insert, Delete, Delete
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:FEED
<Trace>
Selects a trace which provides the points for the power loss list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Setting parameters:
<Trace>
String parameter for the trace name, e.g. '
Trc1
'. The trace must
exist in the active recall set.
*RST:
n/a
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:INSert<ListNo>
<Frequency>, <TransCoeff>
Inserts a new frequency and transmission coefficient at position no. <ListNo> in the
power loss list. The following points are shifted down in the list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo>
Number of point to insert in the list.
Parameters:
<Frequency>
Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range:
Stimulus values outside the frequency range of the
analyzer are allowed.
*RST:
n/a
Default unit: Hz
SCPI Command Reference