Command Reference
R&S
®
ZNA
1262
User Manual 1178.6462.02 ─ 12
Parameters:
<Frequency>
Stimulus frequency value. If several points with identical fre-
quencies are added, the analyzer automatically ensures a fre-
quency spacing of 1 Hz.
Range:
Stimulus values outside the frequency range of the
analyzer are allowed.
*RST:
n/a
Default unit: Hz
<TransCoeff>
Transmission coefficient
Range:
-300 dB to +200 dB
*RST:
n/a
Default unit: dB
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Manual operation:
See
"Frequency / Transm. Coefficients: Insert, Delete, Delete
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>:ALL
Clears the power loss list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
<ListNo>
Number of point in the list. This suffix is ignored; the command
deletes all points.
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Event
Manual operation:
See
"Frequency / Transm. Coefficients: Insert, Delete, Delete
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:DELete<ListNo>[:
DUMMy]
Deletes a single point no.
<ListNo>
in the power loss list.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
SCPI Command Reference