Command Reference
R&S
®
ZNA
1253
User Manual 1178.6462.02 ─ 12
*RST:
RRAFter
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
"Source Leveling/Src. Lev. Method"
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:RRECeiver
<arg0>
Enables or disables calibration of the reference receiver together with the source
power calibration.
Note:
It is not possible to disable flatness calibration
(
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:FLATness
) and the refer-
ence receiver calibration simultaneously.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the verifi-
cation is either enabled or disabled for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the verifica-
tion is either enabled or disabled for all sources.
Parameters:
<arg0>
Enables (ON) or disables (OFF) reference receiver calibration.
*RST:
ON
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Gets the number of available power calibrations (both source power and measurement
receiver).
Use
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
to retrieve the
related settings.
Suffix:
<Ch>
.
Number of the calibrated channel.
<PhyPt>
This suffix is ignored.
<Wv>
This suffix is ignored.
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:DEFault
<arg0>
Enables the analyzer to create a set of default source power calibration data. The ana-
lyzer uses the reference channel power to acquire the default calibration data. No
external power meter is required.
Tip:
SCPI Command Reference