Command Reference
R&S
®
ZNA
1219
User Manual 1178.6462.02 ─ 12
<arg1>
FUNDamental | FIXed
FIXed – use a fixed power, to be specified via
FREQuency:CONVersion:MIXer:PMFixed
FUNDamental – use the channel base power specified via
SOURce<Ch>:POWer<Pt>:STARt, SOURce<Ch>:
POWer<Pt>:STOP
,
SOURce<Ch>:POWer<Pt>[:LEVel][:IMMediate][:
AMPlitude]
*RST:
Minimum of the analyzer's frequency range
Example:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
Manual operation:
See
SOURce<Ch>:FREQuency<PhyPt>[:CW]
<FixedFreq>
SOURce<Ch>:FREQuency<PhyPt>:FIXed
<FixedFreq>
Defines the fixed (Continuous Wave, CW) frequency for all sweep types operating at
fixed frequency ("Power", "Time", "CW Mode"). The two command forms
SOURce<Ch>:FREQuency<PhyPt>:CW
and
SOURce<Ch>:FREQuency<PhyPt>:FIXed
are equivalent.
The frequency range depends on the R&S
.
Note:
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
is equivalent to
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
. Source and receiver frequency are
always equal; the four commands overwrite each other.
With option R&S
ZNA-K4, Frequency Conversion Measurements, port-specific source
and receiver frequencies can be defined; see
CONVersion:ARBitrary:IFRequency
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports in the active
channel.
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency.
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
SCPI Command Reference