Command Reference
R&S
®
ZNA
1190
User Manual 1178.6462.02 ─ 12
Example:
*RST; :SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus NORMal selectivity.
SEGM:BWID:SEL HIGH
Change the selectivity to HIGH.
BWID:SEL?
Query the (default) selectivity for unsegmented sweeps. The
response is NORM.
SEGM:BWID:SEL:CONT OFF
Couple the selectivities in all segments and reset the selectivity
in segment no. 1 to the unsegmented value NORMal.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:DEFine
<StartFreq>, <StopFreq>, <Points>,
<Power>, <SegmentTime>|<MeasDelay>, <Unused>, <MeasBandwidth>[,
<LO>, <Selectivity>, <FreqSweepMode>]
Creates or re-defines a sweep segment no.
<Seg>
with specific channel settings.
Entry of the first seven numeric parameters is mandatory; no default values are provi-
ded. All settings except
<LO>
can be changed for existing segments using other com-
mands of the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
Note:
to create a segment with default
channel settings. Use
[SENSe<Ch>:]SEGMent<Seg>:INSert
new segment into the current segment list.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number. Segment numbers must be sequential.
The specified segment number must be smaller or equal to the
number of existing segments plus 1.
If segment number
<Seg>
already exists, it is replaced by the
new segment.
Parameters:
<StartFreq>
Start frequency of the segment; see
Default unit: Hz
<StopFreq>
Stop frequency of the segment; see
Default unit: Hz
<Points>
Number of sweep points in the segment. See
<Power>
Internal source power in the segment. See
Default unit: dBm
SCPI Command Reference